<del id="6iqyq"></del><li id="6iqyq"><tr id="6iqyq"><cite id="6iqyq"></cite></tr></li>
<th id="6iqyq"><track id="6iqyq"><dl id="6iqyq"></dl></track></th>
  • <rp id="6iqyq"><object id="6iqyq"><input id="6iqyq"></input></object></rp>

  • <rp id="6iqyq"></rp>

    <em id="6iqyq"></em>

          <th id="6iqyq"><track id="6iqyq"></track></th>
          <li id="6iqyq"><acronym id="6iqyq"><u id="6iqyq"></u></acronym></li>
          <form id="6iqyq"></form>
          <em id="6iqyq"><acronym id="6iqyq"></acronym></em>
          <form id="6iqyq"><tr id="6iqyq"></tr></form>
        1. <rp id="6iqyq"></rp>
          <dd id="6iqyq"><track id="6iqyq"></track></dd>

            <tbody id="6iqyq"><center id="6iqyq"><video id="6iqyq"></video></center></tbody>
            1. <span id="6iqyq"></span>
              <th id="6iqyq"></th>
              <nav id="6iqyq"><center id="6iqyq"><table id="6iqyq"></table></center></nav>
              歡迎來到南京延津醫療設備維修中心!
              產品搜索
              在線客服
              • 客服一:

              • 客服二:
              聯系我們

              聯系人: 趙經理

              ??服務熱線:15298381148

              公司電話:025-85305900

              公司傳真:025-57718737

              郵 箱: yanjinkj@163.com

              地 址: ??南京市浦口柳州東路3號天潤城第七街區049

              新聞動態
              美國GE Prospeed Ai螺旋CT維修
              發布時間:2017-02-11   點擊率:2396
               

              高壓部分故障1例

                 故障現象:CT機在掃描病人的過程中出現幾次XG Error,錯誤信息如下:

                  XG error

                  Phase is 5 code is 60

                  Detail error code:60—0313H

                  Exposure Error:ILR max current detected

              最后機器無法掃描。

                  故障排除:查手冊,可能出現故障的原因有:(1)Tube;(2)KV control;(3)HV Tank。做如下診斷:No load HV diagnostic without Anode Rotation nor Filament Heating(即沒有陽極旋轉也沒有燈絲加熱的無負載高壓診斷),打開前蓋,左右側蓋,關閉“R0tate”,“XG Power”,“Slip Ring 115V”,打開SUB板上的Services with ON,用手旋轉Gantry使球管到3點鐘位置,取下HV Tank側高壓電纜,在高壓插座內貯滿高壓變壓器油,把Services with打到0ff位置,打開“XG Power”,“Slip Ring 115V”,單擊顯示器上的SER—VICE圖標,單擊System func Check,單擊Generator Test,單擊KV Diagnostic(no load KV),選擇默認條件曝光,OC上報告錯誤:

                  XG error

                  Phase is 2 code is 50

                  Detail error code:50—0204H

                  Heater error:TGP error detected

                  說明燈絲開路,由此證明:HV Tank和KV control正常,球管故障,關閉機器,等待幾分鐘后拆下球管,發現球管內有5 cm直徑大小的空氣泡,至此說明在曝光時,由于氣泡而高壓放電產生的過電流現象,更換球管,故障排除。